DX-150 Gauss Meter

DX-150  Full-Digital Gauss/Tesla Meter is suitable for testing AC and DC magnetic fields. Maximum range is 30KG (3T), full menu operation, automatic zero calibration, automatic range, 43⁄4 digit reading, high resolution 1/60000, so that you can view 0.1Gs weak changes in the high magnetic field measurement; It is mainly sold to research institutes and various types of production units.

 

DX-150 Full-Digital Gauss/Tesla Meter is designed for applications requiring high precision, high-resolution magnetic field measurement. It is an intelligent solution against the traditional analog, digital table header type Gauss / Tesla meter. It is a high precision intelligent apparatus, which integrates measuring, storage, industrial control, communication just in one machine. DX-150 takes advantage of microprocessor technology to provide you with high-precision, higher refresh rate and display resolution and high measurement repeatability and stability in your research and production.With simple and convenient operation panel and the full menu, 4 ¾-bit resolution readings, a high-resolution 1 / 60000, so that you can view 0.1Gs weak changes in the high magnetic field measurement. It has a maximum range up to 30KG (3T). It has a large capacity of data storage and you can review data according to storage time. In one word, it is suitable for a full range of DC magnetic field measurements, and it is mainly sold to research institutes and various types of production units.

Features

4 ¾-bit resolution on wide-screen LCD panel Auto-zero, automatic/manual range
Maximum/ minimum/ interface locking RS-232C interface
Display units optional Gauss or Tesla Threshold setting and alarming
Probe auto calibration Dozens of optional basic or customized probes
Communication baud rate adjustment Standard 1.00mm ultra-thin consistency probe
Data-storage(auto/manual)/Historical-data reviewing Time setting/ Temperature display

Specifications

Model-150 Specifications

 

0.1mT=1Gs

Range

(0.1Gs-3000Gs)/( 1Gs-30KG)

Resolution

0.1Gs(0.01mT)

Accuracy (display value)

DC

±0.3% of reading ±0.05% of range

AC

±2.0% of reading ±0.5% of range

Frequency Range

DC mode

DC

AC mode

AC20Hz to 20kHz

Additional effects

Temperature coefficient

±(0.03% ±1 count)/ ºC

Temperature range

Working temperature

0ºC to 50ºC

Storage temperature

-10ºC to 60ºC

Panel display

128 * 64 pixel, widescreen LCD

Communication Interface

RS-232   Standard 9-pin D-type connectors

Supply power

Voltage: 200/240v  Frequency: 50-60 Hz

Peripheral: computer, oscilloscope, industrial equipment, bus system

Using occasions

Permanent magnet material surface magnetic field, DC motors, speakers, magnetic separator, and permanent magnetic separator magnetic field of work. Magnetic circuit gap and slit field within the magnetic field. Also it can be applied to the detection machining residual magnetic field of the goods, distribution of magnetic field, the remaining magnetic after demagnetization, the magnetic of motor and magnetic flux leakage of electromagnetic fields and other kinds of magnetic field intensity of magnetic induction.

Main Function

*High –precision magnetic field measurement

* Automatic / Manual Storage: This machine provides 500 data storage space and automatically store data according to set time (1-60 seconds). Also it provides random storage and can be reviewed as per time and stored order.

*Max / Min Lock: Turn on or off the peak measurement function.Capture and display the highest and the lowest magnetic field readings, and lock the interface.
First gear : Maximum/minimum holding 
Second gear: interface locking
Third gear: relieving

*Auto zeroing mode 
1)relative zero magnetic field mode: put the probe into chamber of Zero-Gauss calibrator, zeroing and store the data.
2)eliminate geomagnetic field mode: horizontally put the probe make sure it isto be vertical against geomagnetic field, zeroing and store the data.

*Communications and drawing:
The RS-232C interface function
1) the user can use the computer to read test data, draw various survey map with professional mapping software, and transmit and print chart
2) with serial real-time transmission function of real time monitoring real-time data in a computer
3) query real-time data through fixed instruction

DX-150 optional accessories

Remote control system
Gauss meter professional mapping software
Zero Gauss chamber
Industrial control output components

teslameter-2-2

teslameter-2-3

Probe selection:

gaussmeter-probe

Model L(cm)
approx
T(mm)
max
W
(mm)
A
(mm)
Active
area
 (mm)
Frequency
range
Full Scale
(Standard)
Magnetic 
linearity
(%rdg)
Operating
temp range
Zero Temp
coefficient
(max)
Temp
coefficient
(max)
Stem
material
MCDX800F 6/8/10 1 2.6 1±0.1 0.15 DC-30kHZ 30kG ±0.2%to30kG  -10°Ct0+100°C ±0.08 G/ºC  ±0.03%/C° Brass
MCDX801F 6/8/10 0.8 2.2 1±0.1 0.15 DC-30kHZ 30kG ±0.2%to30kG -10°Ct0+100°C ±0.08 G/ºC  ±0.03%/C° Brass 
MCDX802F 6/8/10 0.6 1.55 1±0.1 0.15 DC-30kHZ 30kG ±0.2%to30kG -10°Ct0+100°C ±0.08 G/ºC  ±0.03%/C° Brass 
MMDX801F 6/8/10 0.5 1.7 1±0.1 0.15 DC-30kHZ 30kG ±0.2%to30kG -10°Ct0+100°C ±0.08 G/ºC  ±0.03%/C° Brass
MMDX802F
Flexible
6/8/10 0.35 1.2 0.5±0.1 0.15 DC-50kHZ 30kG ±0.2%to30kG -10°Ct0+100°C ±0.08 G/ºC  ±0.03%/C° Resin 
MMDX803F
Flexible
6/8/10 0.25 0.7 0.5±0.1 0.1 DC-50kHZ 30kG ±0.2%to30kG -10°Ct0+100°C ±0.08 G/ºC  ±0.03%/C° Resin

Upgradeable to high temperature probe (-10°C - +195°C), low temperature probe (-269°C - +75°C), high and low temperature resistant probe (-196°C - 195°C) and temperature compensated type ( Temperature coefficient: ±0.01%/°C).

High-Precision-Axis-Probes

Model L(cm)
approx
D(mm)
max
A(mm) Active area
(mm)
Frequency
range
Full Scale
(Standard)
Magnetic
linearity
(%rdg)
operating
temp range
Zero Temp
coefficient
(max)
Temp
coefficient
(max)
Stem
material
1ADX800F 10/25/50 5 1±0.1 0.15 DC-30KHZ 30kG ±0.2%to30kG  -10°Ct0+100°C ±0.08 G/ºC   ±0.03%/C° Brass
1ADX800F 10/25/50 3 1±0.1 0.15 DC-30KHZ 30kG ±0.2%to30kG -10°Ct0+100°C ±0.08 G/ºC  ±0.03%/C° Brass 
1ADX802F 10/25/50 2 1±0.1 0.15 DC-30KHZ 30kG ±0.2%to30kG -10°Ct0+100°C ±0.08 G/ºC  ±0.03%/C° Brass 
1ADX803F 10/25/50 1.5 1±0.1 0.15 DC-30KHZ 30kG ±0.2%to30kG -10°Ct0+100°C ±0.08 G/ºC  ±0.03%/C° Brass
1ADX804F 10/25/50 1.2 0.5±0.1 0.15 DC-30KHZ 30kG ±0.2%to30kG -10°Ct0+100°C ±0.08 G/ºC  ±0.03%/C° Brass 
1ADX63805F 10/25/50 0.9 0.5±0.1 0.1 DC-30KHZ 30kG ±0.5%to30kG -10°Ct0+100°C ±0.08 G/ºC  ±0.03%/C° Brass

Upgradeable to high temperature probe (-10°C - +195°C), low temperature probe (-269°C - +75°C), high and low temperature resistant probe (-196°C - 195°C) and temperature compensated type ( Temperature coefficient: ±0.01%/°C).

Flexible-High-Precision-Transverse-Probes

Model L(cm)
approx
T(mm)
max
W
(mm)
A
(mm)
Active
area
 (mm)
Frequency
range
Full Scale
(Standard)
Magnetic 
linearity
(%rdg)
Operating
temp range
Zero Temp
coefficient
(max)
Temp
coefficient
(max)
Stem
material
CDX800F-R 6/8/10 1.5 2.4 1±0.1 0.15 DC-50KHZ 30kG ±0.2%to30kG  -10°Cto+70°C    ±0.08 G/ºC   ±0.03%/C° plastic
CDX801F-R 6/8/10 1 2.4 1±0.1 0.15 DC-50KHZ 30kG ±0.2%to30kG -10°Cto+70°C ±0.08 G/ºC  ±0.03%/C° plastic
CDX802F-R 6/8/10 0.8 2.3 1±0.1 0.15 DC-50KHZ 30kG ±0.2%to30kG -10°Cto+70°C ±0.08 G/ºC  ±0.03%/C° plastic

 

 

……

Home    Gauss Meter    DX-150 Gauss Meter
Magnetizer and Demagnetizer
Power Current Source
Hall Probe
Other Supporting Instruments
Permanent Magnet Testing
Hall Effect System
Testing Systems
Solenoid
Helmholtz Coil
Electromagnet
Magnetic Field Generator
Flux-gate Meter
Flux Meter
Gauss Meter